Rheometer – Stability against shear

Stability against shear

  • DIN 51810-1
  • For Oils, greases and pastes
  • Focuses on the total system
  • Rheology is the measuring and description of the flow properties of a substance and how its behavior is influenced by solid and liquid components.
  • Relevant for pumps, conveying, stirring, mixing and bearing applications
  • Customizable in the test setup for rotating and oscillating shear, in temperature and mechanical load
  • →Test setup can be customized to meet the precise problem question
  • Describes dynamic viscosity and mechanical stability before and after shear

Tribo-cell

  • Tibolo-cell-PTD 200 (Anton Paar)
  • For Oils. Greases, pastes, dispersion, bonded coatings
  • Characterization of the lubricity (coefficient of friction and wear) for the model system ball-on-three-plates udner mechanical load
  • Load: 1-23 N normal force
  • Temperature range: -40 - +200°C
  • Speed: up to 3000 U/min
  • Determination of the Stribeck graph and measurement of the coefficient of friction
  • Bearing option for testing the lubricants lubricity in bearings